Comparison of Two Clinical Chemistry Analyzers by Total Analytical Error and Measurement Uncertainty
xmlui.dri2xhtml.METS-1.0.item-rights
info:eu-repo/semantics/closedAccessDate
2023Metadata
Show full item recordAbstract
Background: Although analytical errors contain a small portion of laboratory errors, they are important in terms of intervention ability and practicality of follow-up by laboratory professionals. Also, from this point of view, the test results' quality, reliability, and accuracy are crucial to laboratories. Therefore, to determine analytical performance parameters for quality management in the analytical phase, clinical laboratories utilize total analytical error (TAE), bias, coefficient of variation (CV), and uncertainty of measurement (MU). Methods: Fifteen biochemistry parameters were compared with Beckman Coulter AU 5800 for 2017 - 2018 and Roche Cobas 8000 for 2019 - 2020 in terms of TAE and MU. The results were evaluated between devices and compared with the EuBIVAS, CLIA, RCPA, PRDEQA%, pUQEAS%, pU%, and TEa-TR datasets. Results: There were no significant differences between the devices for the mentioned periods. Device performances resulted in similar outcomes. During our four-year study, nearly all of our tests failed for EuBIVAS, RCPA, and pU%. On the contrary, almost all of our parameters gave valid results according to the CLIA, PRDEQA%, pUQEAS%, and TEa-TR ranges. Conclusions: It is crucial to distinguish between mistake and uncertainty. The discrepancy between the measured value and the 'actual value' is called error. Uncertainty is a measure of how confident you are in the measurement outcome. We endeavor to remedy any known inaccuracies wherever feasible by applying adjustments from calibration certifications. On the other hand, any inaccuracy whose value is unknown is a cause of doubt.